TX 2000
TXRF Analysis
Laboratory TXRF spectrometer for quantitative multi-element trace and ultra-trace analysis, with automatic primary-beam switching.

About the TX 2000
An outstanding spectrometer that allows to perform Energy Dispersive X-Ray Fluorescence Analysis in Total Reflection Geometry.
TX 2000 is the state of the art laboratory spectrometer for quantitative multi element trace analysis using Total Reflection X-Ray Fluorescence (TXRF).
TXRF is a type of Energy Dispersive X-Ray Fluorescence (EDXRF) in which the beam strikes the sample, deposited as a thin layer on a carrier, at a very small incidence angle, in order to exploit enhancement given by the total reflection effect. As far as the sample is thin enough (thin film approximation), the tecnique’s main strenghts are:
TX 2000, equipped with an automatic primary beam switching system (MoKa, WLa/Lb and bremsstrahlung 33 keV), a Silicon Drift Detector (SDD) and an high power X-Ray Tube is the optimal solution for applications in environmental, chemicals and nuclear fields.
- simultaneous multi-element analysis from Na to Pu
- no matrix effect
- no matrix-dependent calibration curves
- improved detection limits (LOD) down to ng/g or lower
- minimum amount of sample
- microanalysis capabilities
- Stepper motors with optical encoders ensure extremely precise angular values.
- Enhanced excitation mode by selection of the most suitable monochromatized energy (W-La/W-Lb/Mo-Ka) provided by the combination of double anode MO/W X-ray tube and W/Si multilayer. Other X-Ray anodes can be monochromatized as well.
- Peltier-cooled Silicon Drift Detector with an energy resolution better than of 133eV (FWHM@MnKa, 1ms peaking time)
- Minimal distance between the sample and the detector (mounted to the axis normal to the plane of the sample).
- Instrumental detection limits below ng/g.
- Helium flux attachment to improve the detection limits for the light elements.
- No matrix effect
Documented applications
- Environmental Analysis: water, dust, sediments, aerosol (ISO/TS 18507:2015, ISO 20289:2018)
- Medical/Pharma: toxic elements in biological fluids and tissue samples, catalyst residues
- Forensic Science: analysis of extremely small sample quantities
- Pure chemicals: acids, bases, salts, solvents, water, ultrapure reagents
- Oils and greases: crude oil, essential oil, fuel oil, gasoline, jet fuel
- Pigments: ink, oil paints, powder
- Nuclear Industry: measurements of radioactive elements
Matrices, elements and ranges depend on the configuration supplied; EMUS confirms them in the quotation.
Sectors served
GNR application fields
Technical data
- Maximum Output Power
- 3 kW
- Max Output Voltage
- 60 kV
- Max Output Current
- 60 mA
- Voltage Step Width
- 0.1 kV
- Current Step Width
- 0.1 mA
- Type
- Glass, Mo/W anode, long fine focus
- Focus
- 0.4 x 12 mm
- Max Output
- 2.5 kW
- (*) Different anodes can be mounted on request
- Type
- W/Si
- Reflectivity 80%
- W-La
- 8.4 keV (ΔE<1 keV)
- W-Lb
- 9.67 keV (ΔE<1 keV)
- Mo-Ka
- 17.44 keV (ΔE<1 keV)
- Bremsstrahlung
- 33 keV
- 12 position sample holder
- Active area
- 40 mm2 as standard. (50 mm2 as option)
- Energy resolution
- ≤ 133 eV FWHM @Mn-Ka (1 μs peaking time)
- Air or He Flux
- Dimensions
- Width 580 mm, Height 1631.5 mm, Depth 793 mm
- Weight
- 200 kg (without chiller)
- Leakage X-Rays
- < 1 mSv/Year (full safety shielding according to the international guidelines)
- External Chiller (1146h, 570w, 740d)
Reproduced from official GNR product information. Values apply to the configuration described by GNR; final configuration is confirmed in the EMUS quotation.
Operating software
- GNR proprietary software for T XRF supports several tasks of analysis, from Data Acquisition, having the full control of all the process and hardware settings (generator and tube, detector, measurements set up) to Data Analysis:
- Least square Marquardt fit procedure for the area calculation (spectral analysis)
- Automatic/manual search function
- Manual or automatic calibration of energy
- Quantification via an internal standard using theoretical and experimental sensitivity curves
Alternatives from other manufacturers
Comparable systems we also represent. If your application requires a different balance of speed, resolution or footprint, these are worth reviewing.
ECO 160 DS
Low weight portable X-ray system designed and built in Denmark using refined EVO technology, with a high-quality metal ceramic X-ray insert protected by a sturdy, IP65-rated casing for full operation in dusty and wet conditions.
CompareComet X-rayECO 200 DS
Low weight portable X-ray system designed and built in Denmark using refined EVO technology, with a high-quality metal ceramic X-ray insert protected by a sturdy, IP65-rated casing for full operation in dusty and wet conditions.
CompareComet X-rayEVO 160D
Robust and reliable portable X-ray system designed and built in Denmark, with a high quality metal ceramic X-ray tube and an IP65-rated composite casing for full operation in dusty and wet conditions.
CompareComet X-rayEVO 200D
Robust and reliable portable X-ray system designed and built in Denmark, with a high quality metal ceramic X-ray tube and an IP65-rated composite casing for full operation in dusty and wet conditions.
CompareHow EMUS supports this
Equipment is only part of the answer. Every system we supply is backed by regional engineering, service and application support.
Mechanical, electrical, environmental setup done right the first time.
Qualification and acceptance testing against real criteria.
Operator and application training built around your parts.
Traceable calibration to keep measurements audit-ready.
Preventive and corrective service across the Balkan region.
Method development and problem-solving with our engineers.
GNR Analytical Instruments
TheTX 2000 is part of the GNR Analytical Instruments portfolio, supplied and supported by EMUS across the Balkan region.
Request Technical Consultation
Talk to an EMUS engineer about theTX 2000, your materials and the analytical configuration that fits your laboratory or production line.